Título: Ventilatory changes during the use of heat and moisture exchangers in patients submitted to mechanical ventilation with support pressure and adjustments in ventilation parameters to compensate for these possible changes: a self-controlled intervention study in humans
Título(s) alternativo(s): Alterações ventilatórias durante o uso de trocadores de calor e umidade em pacientes submetidos à ventilação mecânica com pressão de suporte e ajustes nos parâmetros ventilatórios para compensar estas possíveis alterações: estudo de intervenção autocontrolado em humanos
Autor(es): Lucato, Jeanette Janaina Jaber
Cunha, Thiago Marraccini Nogueira da
Reis, Aline Mela dos
Picanço, Patricia Salerno de Almeida
Barbosa, Renata Cléia Claudino
Liberali, Joyce
Righetti, Renato Fraga
Palavras-chave: Respiração artificial
Umidificadores
Desmame do respirador
Temperatura
Unidades de terapia intensiva
Data do documento: 2017
Editor: Associação de Medicina Intensiva Brasileira - AMIB
Citação: Lucato, Jeanette Janaina Jaber, et al. “Ventilatory changes during the use of heat and moisture exchangers in atients submitted to mechanical ventilation with support pressure and adjustments in ventilation parameters to compensate for these possible changes: a self-controlled intervention study in humans”. Revista Brasileira de Terapia Intensiva, vol. 29, no 2, 2017. DOI.org (Crossref), https://doi.org/10.5935/0103-507X.20170026.
Resumo: Objective: To evaluate the possible changes in tidal volume, minute volume and respiratory rate caused by the use of a heat and moisture exchanger in patients receiving pressure support mechanical ventilation and to quantify the variation in pressure support required to compensate for the effect caused by the heat and moisture exchanger.Methods: Patients under invasive mechanical ventilation in pressure support mode were evaluated using heated humidifiers and heat and moisture exchangers. If the volume found using the heat and moisture exchangers was lower than that found with the heated humidifier, an increase in pressure support was initiated during the use of the heat and moisture exchanger until a pressure support value was obtained that enabled the patient to generate a value close to the initial tidal volume obtained with the heated humidifier. The analysis was performed by means of the paired t test, and incremental values were expressed as percentages of increase required.Results: A total of 26 patients were evaluated. The use of heat and moisture exchangers increased the respiratory rate and reduced the tidal and minute volumes compared with the use of the heated humidifier. Patients required a 38.13% increase in pressure support to maintain previous volumes when using the heat and moisture exchanger.Conclusion: The heat and moisture exchanger changed the tidal and minute volumes and respiratory rate parameters. Pressure support was increased to compensate for these changes.